TOF-SIMS

PHI nanoTOF 3+

Time-of-Flight Secondary Ion Mass Spectrometer (TOF-SIMS)

Pushing the boundaries of nanoscale surface mass spectrometry.

Building on the TRIFT analyzer (Triple Focusing Electrostatic Analyzer) renowned for its superior ion transmission, and incorporating the proven sample transfer mechanism from our XPS systems, PHI nanoTOF 3+ features unique ion beam technology that enables turnkey charge neutralization. It delivers automated multi-sample analysis with low noise and high sensitivity, while offering a wide array of configurable options for virtually any application.

KEY TECHNOLOGY
  • High-Precision Analysis with TRIFT Analyzer
  • Continuous Unattended Measurement of Multiple Insulating Samples
  • Unique Ion Beam Technology
  • Molecular Structure Analysis via Parallel Imaging MS/MS
  • Wide Range of Options
PHI nanoTOF 3+
TRIFT Analyzer

Triple Focusing Electrostatic Analyzer for Complex Sample Geometries

TRIFT Analyzer for a Wide Range of Sample Geometries, Wide Bandpass Energy, and Wide Acquisition Angle

In exactly the same mass will have different flight times in the analyzer. This is particularly noticeable for samples with height differences or uneven surfaces.

The Triple Ion Focusing Time-of-flight (TRIFT) spectrometer is able to simultaneously compensate for differences in flight time caused by differences in energy and emission angle. This is a key feature of the TRIFT spectrometer, which achieves high mass resolution and high detection sensitivity, as well as imaging with reduced shadows as demonstrated in the examples below.

TRIFT Analyzer Configuration Diagram and Micro Ball Measurement Examples
Primary Ion Gun

Primary Ion Gun for High-Precision Analysis

Advanced Ion Beam Technology Achieves High Spatial Resolution and High Mass Resolution

PHI nanoTOF 3+ offers TOF-SIMS analysis with a high spatial resolution of < 500 nm in high mass resolution mode and < 50 nm in high spatial resolution mode.

The combination of a high brightness ion source, a high precision pulse mechanism and a high resolution spectrometer ensures low noise, high sensitivity and high resolution measurements. In both modes, measurements are completed within a few minutes.

High spatial resolution pulsed. beam <50nm PHI nanoTOF3 [conventional] PHI nanoTOF3′
45 nm Spatial Resolution Measurement Example
TOF-SIMS images from medicine
Chemical Imaging Measurement Example by m/z
Unattended Measurement

The first-ever automated unattended TOF-SIMS analysis of multiple insulating samples

With the newly developed auto sample bias adjustment feature, it is now possible to perform unattended TOF-SIMS measurements on samples including insulating samples. The process is thoroughly simplified; 1) taking an intro photo, 2) selecting the analysis points on the intro photo, and 3) pressing the start button. In the past, skilled operators had to be dedicated to the instrument for analysis, but now the high quality data can be obtained regardless of operator skill.

Unattended Multi-Sample Measurement of Insulating Materials (Sample Holder)

Automated Sample Transport System and Parking Position as Standard

The instrument is equipped with an automatic sample transfer system that has been proven in more than 300 XPS Q-Series units. Sample sizes of up to 100 mm × 100 mm are possible, and the analysis chamber has a built-in parking position as standard. The system allows continuous automated unattended analysis of large numbers of samples.

Fully Automated Sample Transfer Stage
Ion Beam Technology

Unique Ion Beam Technology

New Pulsed Argon Ion Gun for Complete Turnkey Charge Neutralization

The pulsed low-energy electrons and argon ions can easily neutralize surface charge for both positive and negative ion polarities regardless of the type and shape of the insulating sample.
* For Ar ion gun option is selected.

Schematic Diagram of Turnkey Charge Neutralization
FIB

FIB Fabrication Followed by Cross-section Imaging within a Single LMIG

FIB Function with Standard LMIG

FIB processing function has newly been added to LMIG for primary ion source. This makes it possible to perform from FIB processing to cross-sectional TOFSIMS imaging with a single ion gun. Stage rotation is involved from processing to observation, but the Eucentric function allows the stage to return to the same field of view after rotation, easy identifying the FIB position.
To achieve 3D tomography, the dedicated Ga-FIB option is recommended. The Ga-FIB may be further optioned to operate as a second analysis beam.

Eucentric Rotation and FIB-TOF imaging of Battery Materials
Parallel Imaging MS/MS

Molecular Structure Analysis based on Parallel Imaging MS/MS (Option)

Simultaneous Acquisition of MS1 and MS2 Data

By combining a tandem mass spectrometer (MS2) with a conventional TOF-SIMS analyzer (MS1), MS1 and MS2 data can be acquired simultaneously in the same measurement field with high sensitivity and speed.
The MS1 spectrum can be complex due to the large number of mass peaks generated by all components and fragments. Conversely, the MS2 spectrum, which is extracted and segmented by the precursor selector, has a much simpler spectral pattern. This provides information about the molecular structure, such as terminal groups, linear chains, or double bonds, and enables specific molecular structure imaging.

Peak Assignment by MS¹ and MS² Spectra (Benzalkonium · Tinuvin 770) MS¹/MS² Spectra Comparison TOF-SIMS Images from polymer surface

Detail of the parallel imaging MS/MS option detail is in the following links.

What is Parallel Imaging MS/MS Option ?
Remote Access

Reliable Remote Access for Measurements and Diagnostics, Even Away from the Instrument

Remote Access Function Allows the Instrument to be Controlled Remotely

The PHI nanoTOF 3+ allows you to access the instrument via your company’s local area network or the internet. Once the sample holder is mounted in the intro chamber, functions such as transferring the sample into the analysis chamber, sample holder exchange with the parking position, setting of the measurement conditions (including all options), and data acquisition can all be performed remotely. Diagnostics of the instrument hardware and electronics can also be accessed remotely.
* For more information on remote diagnostics, please contact our Customer Service Department.

Remote Access Connection Diagram
Options

Optional Accessories

  • Parallel Imaging MS/MS
  • Ar Gas Cluster Ion Beam Gun (GCIB)
  • C60 Ion Gun
  • Cs Ion Gun
  • Ar/ O2 Gas Gun
  • Hot/Cold Sample Stage Module
  • High Temperature Sample Stage Module
  • Transfer Vessel
  • Oxygen Flood Module
  • Focused Ion Beam (FIB) Gun
  • Preparation Chamber
  • Various Special Holders
  • Off-line Data Reduction Software
  • Static SIMS Library